Self-Supervised Mask-Aware Transformers for Fault-Tolerant FBG Force Sensing in Minimally Invasive Surgical Robotics
arXiv:2606.18628v1 Announce Type: new Abstract: In minimally invasive surgical robotics, catheter-scale Fiber Bragg Grating (FBG) sensors are promising due to their ability to estimate multi-dimensional forces by multiplexing several optical channels. However, deploying these compact multi-channel sensors introduces two critical engineering challenges: inherent nonlinear cross-axis coupling during complex deformations, and intermittent channel dropouts caused by fiber fractures in constrained workspaces. These compounding issues severely degrade force estimation. Existing fault-tolerant approaches rely on combinatorial model banks, which scale exponentially with the channel count and demand prohibitively expensive per-pattern calibration. In this paper, we propose a unified, self-supervised mask-aware Transformer that explicitly models channel availability to enable graceful degradation under diverse and dynamic sensor failures. The encoder is pretrained via masked-channel reconstructi